phone flowmeters

Flow Rig Calibration Certificate

flowmeters-homeright.jpg
Model ..................              Calibration Date ................
Serial No. .................. Calibration Due ................
Room Temp. ..................     
  
Run Number
Translator pulses
Liquid Temp.
Translator Temp.
Draw Pressure
Weight (grams)
Corrected Weight
1
1000
18.54
18.54
60
364.50
363.96
2
1000
18.54
18.54
60
364.50
363.96
3
1000
18.54
18.54
60
364.60
364.06
4
1000
18.55
18.55
60
364.50
363.96
5
1000
18.57
18.57
60
364.45
363.91
6
1000
18.57
18.57
60
364.45
363.91
7
1000
18.57
18.57
60
364.35
363.81
8
1000
18.59
18.59
60
364.50
363.96
9
1000
18.59
18.59
60
364.45
363.91
10
1000
18.65
18.65
60
364.55
364.01
Average
1000
18.571
18.571
60
364.49
363.95
       
Water density @ temp
0.9985309
     Weight max
364.60
364.06
Water volume
365.02
Weight min
364.35
363.81
     Error ±%
0.03429496
Correction for compressibility
  
  
  
  
1.000115         
Correction for thermal expansion of tube
0.9999542
Correction for thermal expansion of translator
0.999983
Correction for expansion of tube for pressure
1.000052
       
Rig calibration constant     2743.883 Pulses/litre  
       
Scales used:-     Jadever Snug - 1500                    
Serial number :- 4034312
Recalibration due :- ............
   
Calibrated by ................. ................ Date ..............